Sample Preparation
Preparation Workflow
Parameters & T-ESPRIT Calculation
1. Sample & Diluent
Daily Liquid Sample Logbook
Streamlined per-day liquid analysis register linking sample matrices, dilution recipes, Bruker T-ESPRIT internal standard entries, and measured excitation tubes.
Analysis Day Selector & Batch Controls
Pick a single analysis date, browse logged days, or batch-register multiple liquid samples using your active Tab 1 preparation recipe.
Logged Liquid Samples
Inline-editable register. Click Process → on any sample to upload Bruker .xlsx / .spx files and run QA/QC.
| Sample Name / ID | Date | Matrix | DF | T-ESPRIT Internal Standard(s) | Tubes Measured | Bruker Files | Notes | Actions |
|---|
Data Processing, 3-Tube Protocol & QA/QC Engine
Attach Bruker S4 T-STAR liquid quantification exports (.xlsx / .spx) across
Mo-K → W-Brems → W-L. Internal-standard quantification eliminates geometric Correction Factors automatically.
Step 1 · Sample Selection & Bruker 3-Tube Import
Strict Bruker excitation order: Mo-K, W-Brems, W-L. No geometric Correction Factors required for liquid ISTD analysis.
.xlsx, .csv, or .spx files here
— excitation tube (Mo-K, W-Brems, W-L) is auto-detected from filename, XML metadata, or K-edge physics.
Step 2 · Bruker S4 T-STAR Optimum Measurement Protocol (Z = 1..103)
Click any element to cycle condition (Mo-K K → Mo-K L → W-Brems K → W-L K → ISTD → Excluded). Detected elements show a bright white inner border.
Step 3 · Liquid Concentrations & QA/QC Engine
Identical T-AERO QA/QC rigor adapted to liquid concentration units (mg/L, µg/L, ng/L) with automatic unit normalization.
| Element | Z | Tube / Line | Conc (mg/L) | ± 1σ (mg/L) | Rel. Err % | LOD (mg/L) | LOQ (3×LOD) | Net Counts | Background | QC Status | Override |
|---|
Raw Bruker .SPX Spectrum Inspector
—
Scientific Documentation & Analytical Methodology
Liquid TXRF internal-standard quantification equations, dilution correction proof, multi-ISTD excitation coverage, and Bruker S4 T-STAR 3-Tube QA/QC criteria.
1. Liquid TXRF Internal-Standard Quantification & Dilution Correction
Why dividing by original sample volume (Vsample,orig) makes Bruker T-ESPRIT automatically correct for dilution, and why drop volume cancels.
In Total Reflection X-Ray Fluorescence (TXRF) of thin dried liquid films on 30 mm quartz reflectors, matrix absorption and secondary enhancement effects are negligible. Quantification is governed solely by the relative fluorescence intensity ratio between analyte i and a homogeneously spiked Internal Standard (IS):
Inside the mixing Eppendorf vial, the total mass of internal standard added is mIS = Cstock × VIS, while the mass of analyte i present comes exclusively from the original undiluted sample aliquot: mi = Ci, orig × Vsample, orig.
When a small drop (5 µL or 10 µL) is pipetted onto the 30 mm quartz carrier and dried, both analyte and internal standard are transferred in the exact same fraction f = Vdrop / Vtotal. Because f appears identically in numerator and denominator, both total vial volume (Vtotal) and reflector drop volume (Vdrop) cancel out identically:
For two-step (serial) pre-dilutions with pre-dilution factor Dpre, substitute the effective undiluted sample volume Vsample, orig, eff = Valiquot / Dpre.
| Lab Verification Case (Hagen 100 mg/L = 0.1 g/L Ga Stock) | Vsample, orig | Vdiluent (H₂O) | VIS (Ga) | Dilution Factor | T-ESPRIT Standard Conc. |
|---|---|---|---|---|---|
| Case A · Undiluted (1×) | 1000 µL | 0 µL | 10 µL | 1.01× | 1.0 mg/L |
| Case B · 10× Dilution | 100 µL | 890 µL | 10 µL | 10.0× | 10.0 mg/L |
| Case C · 100× Dilution | 10 µL | 980 µL | 10 µL | 100.0× | 100.0 mg/L |
2. Multi-Internal-Standard Strategy Across Bruker S4 T-STAR Tubes
Selecting interference-free internal standards for full elemental coverage across W-L, Mo-K, and W-Brems excitations.
Fe–Y, Z=26..39) and heavy L-lines (Ce–U, including Pb-L, Hg-L).
Ga (Z=31) — rare in environmental liquids, crisp Kα at 9.25 keV. Alternatives: Y, Se, Co.Zr–I, Z=40..53, including Cd-K, Ag-K, Sn-K, Sb-K) above the Mo K-edge.
Pd (Z=46) — clean Kα at 21.18 keV. Alternative: In or Y.Na–Mn, Z=11..25) with superior sensitivity for Na, Mg, Al, Si, P, S, Cl, K, Ca, Ti, Cr, Mn.
Sc (Z=21) — Kα at 4.09 keV, cleanly separated from Ca Kβ/Ti Kα. Alternative: V.3. Bruker 3-Tube Optimum Protocol & QA/QC Acceptance Criteria
Automated quality control rules ensuring publication-grade liquid TXRF elemental concentrations.
- Maximum Relative Error (≤ 33.3%): Elements whose counting uncertainty exceeds 33.3% (
σ / C × 100% > 33.3%, i.e. SNR < 3) are automatically flagged and filtered. Configurable for Strict (≤ 20%) or Relaxed (≤ 50%) studies. - Background Dominance Filter (Bkg ≤ Net): Any peak where spectral background exceeds net peak intensity (
Background > Net) is rejected to prevent false positives near trace edges. - Zero-Count & Numeric Integrity (Net > 0): Zero-count or "Not determined" rows from T-ESPRIT are strictly excluded from reported summaries.
- Optional LOQ Screening (C > 3 × LOD): Enables strict Limit-of-Quantification screening where only analytes exceeding
3 × LLDpass QA/QC. - Internal Standard Exclusion & QC Callout: Elements tagged as
ISTDin the Periodic Table (defaultSc, Ga, Pdplus any custom IS synced from Tab 1) are excluded from the analyte table and verified separately in the ISTD QC Check panel. - Bruker
.spxXML Field Mapping:Net = NetIntens,Error = round(NetIntens × Sigma),Background = Background,LifeTime_s = LifeTime / 1000.